Application

Permanent magnet for MRI of the head and extremities

Contributors

Martin Häuer1
Tom O’Reilly2
Sebastian Schachel1
Wouter Teeuwisse2
Lukas Winter1

Estimated cost

10.000€

Progress

License: CERN OHL W v2

System Specifications

  • Permanent magnet in “ROMA” (Rotation Optimized Magnet) array, which differs from previously used Halbach-based magnet designs (v1*) in that the orientation of the individual magnets inside the array no longer follow the strict Halbach definition but rather are optimized to maximize homogeneity
  • 1936× (12×12×12)mm³ NdFeBr N52 magnets
  • 384× (50×12×12)mm³ NdFeBr N52 magnets
  • B0 ~ 50mT
  • Weight ~100kg
  • Bore = ø322mm × 428+14mm (with back plate, without gradients)
  • outer dimensions (l×b×h) = (450×520×567)mm³ (with back plate, without gradients)

Publications

Affiliations

1Physikalisch-Technische Bundesanstalt, Berlin, Germany
2Leiden University Medical Center (LUMC), Leiden, Netherlands

Figures